Performance Data

Flow Measurement Uncertainty
Standard: 0.20 % of rate
Enhanced: 0.15 % of rate
Premium: 0.10 % of rate
Ultimate: 0.05 % of rate – requires RHE 40 Series transmitter
Low Flow Calibration: 0.10 % of rate – requires RHE 40 Series transmitter
Customized Calibration: Consult factory

Low flow calibration focuses on the range from ½ × Qₙₒₘ downwards to lower flow rates than on other calibrations. Often used for low pressure gas or very viscous liquids. Customized calibration uses specific calibration points according to customer requirements.

Flow Measurement Repeatability
Standard Sensors: ±0.1 % of rate
Premium Sensors: ±0.05 % of rate

Temperature Measurement
Better than ±1 °C

Density Calibration (Liquid)
Standard: ±0.0035 kg/l uncertainty
Premium: ±0.0005 kg/l uncertainty – requires High Accuracy or Focused Calibration Option

Reference conditions: 18–24 °C Water @ 1–3 bar

Measurement Tube Pressure Ratings

The maximum pressure (pₘₐₓ) of a sensor is determined by its lowest rated part: the measurement tube (see table below), the connection block (see datasheet) or the process connection (see corresponding standards or manufacturer information).

If applicable, the material of a connection block is always SS 316L. For sensors in sealless versions, the material of the process connection is usually the same as the material of the measurement tubes. In special cases, the material may differ.

 

50 °C
122 °F

120 °C
248 °F

210 °C
410 °F

350 °C
662 °F

SS 316L (standard)

452 bar
6556 psi

402 bar
5831 psi

339 bar
4917 psi

280 bar
4061 psi

Alloy C22

474 bar
6875 psi

421 bar
6106 psi

358 bar
5192 psi

297 bar
4307 psi

Super Duplex *
Low Pressure

630 bar
9137 psi

540 bar
7832 psi

410 bar
5947 psi

 

Super Duplex *
Medium Pressure

1050 bar
15229 psi

900 bar
13053 psi

720 bar
10443 psi

 

Super Duplex *
High Pressure

1481 bar
21480 psi

1379 bar
20001 psi

1236 bar
17927 psi

 

Sandvik HP160®

1219 bar
17680 psi

1175 bar
17042 psi

795 bar
11531 psi

 

Tantalum

257 bar
3727 psi

198 bar
2871 psi

161 bar
2335 psi

 

* Minimum operating temperature for Super Duplex stainless steel is -40 °C. Maximum pressure may be lower depending on the design type.

Uncertainties and flow measurement turn-down

The turn down capability from Qₙₒₘ of the flow sensor is driven mainly by its zero point stability. At the very low end of the measuring range the uncertainty (u) is dominated by zero point stability.

• Zero point stability of a standard sensor with general accuracy calibration: 0.000036 kg/min (0.000079 lbs/min).
• Zero point stability of a premium sensor with high accuracy or focused calibration: 0.000019 kg/min (0.000042 lbs/min).
• uncertainty = calibration uncertainty, if: flow ≥ (zero point stability / (base uncertainty/100))
• uncertainty = (zero point stability/Q)*100, if: flow < (zero point stability / (base uncertainty/100))

Uncertainties from environmental and process conditions

If sensors are not zeroed at operating conditions, minor additional uncertainties can arise from elevated temperatures and pressures:

• ±0.00308% of maximum flow per °C
• ±0.0208% of maximum flow per bar

Process temperature effect on density:
• Additional uncertainty of ±0.000641 g/cm³ per °C difference from calibration temperature with standard density calibration.
• Additional uncertainty of ±0.000073 g/cm³ per °C difference from calibration temperature with enhanced density calibration.
• This effect can be mitigated by a simple field density adjustment at operating conditions.

Process pressure effect on mass flow:
• The effect of pressure on flow measurement is 0.001% of rate per bar. Compensation is possible by pressure sensor input (analog input or digital write) or manual value entry into the transmitter.

Process pressure effect on density:
• The effect of pressure on density measurement is 0.00012 g/cm³ per bar. Compensation is possible by pressure sensor input (analog input or digital write) or manual value entry into the transmitter.

Applying Premium calibration to higher pressure, special materials and/or non-standard temperature models may show higher zero uncertainties (up to 3 times higher than standard sensor).

Couldn’t find what you were looking for?

Contact us with your specification for assessment and availability of other sensor configurations.

Other wetted materials may be possible for chemical compatibility, lower pressure drop, abrasion allowance or other application specific requirements. Rheonik can provide nearly any material for the wetted parts.