RHM 15

FLOW SENSOR

The RHM 15 is an all-purpose meter suitable for liquid processing plants. A wide application flow range capability allows for commonality of installation across numerous lines – a real benefit to owners and operators.

  • Range: 0.5 kg/min up to 240 kg/min
  • Pressure: up to 1282 bar (18593 psi)
  • Temperature: -196°C (-321°F) up to +350°C (662°F)
  • Accuracy: 0.1 % of rate
  • Materials: SS 316L / SS 316Ti, Alloy C22 – 2.4602, Super Duplex – 1.4410, Tantalum – UNS R05200
  • Process Connection: ¾" up to 2 ½"
CONFIGURE
CATEGORY Medium Size Sensors, Chemical Injection, Fuel Consumption, Oil Measurement, Painting/Coating Lines, Plastics Production

Performance Data

Performance Data

Flow Measurement Uncertainty
Standard: 0.20 % of rate
Enhanced: 0.15 % of rate
Premium: 0.10 % of rate
Ultimate: 0.05 % of rate – requires RHE 40 Series transmitter
Low Flow Calibration: 0.10 % of rate – requires RHE 40 Series transmitter
Customized Calibration: Consult factory

Low flow calibration focuses on the range from ½ × Qₙₒₘ downwards to lower flow rates than on other calibrations. Often used for low pressure gas or very viscous liquids. Customized calibration uses specific calibration points according to customer requirements.

Flow Measurement Repeatability
Standard Sensors: ±0.1 % of rate
Premium Sensors: ±0.05 % of rate

Temperature Measurement
Better than ±1 °C

Density Calibration (Liquid)
Standard: ±0.001 kg/l uncertainty
Premium: ±0.0005 kg/l uncertainty – requires High Accuracy or Focused Calibration Option

Reference conditions: 18–24 °C Water @ 1–3 bar

Measurement Tube Pressure Ratings
Measurement Tube Pressure Ratings

The maximum pressure (pₘₐₓ) of a sensor is determined by its lowest rated part: the measurement tube (see table below), the connection block (see datasheet) or the process connection (see corresponding standards or manufacturer information).

If applicable, the material of a connection block is always SS 316L. For sensors in sealless versions, the material of the process connection is usually the same as the material of the measurement tubes. In special cases, the material may differ.

 

50 °C
122 °F

120 °C
248 °F

210 °C
410 °F

350 °C
662 °F

SS 316Ti (standard)
Low Pressure

161 bar
2335 psi

149 bar
2161 psi

135 bar
1958 psi

117 bar
1696 psi

SS 316Ti (standard)
High Pressure

389 bar
5642 psi

359 bar
5206 psi

326 bar
4728 psi

283 bar
4105 psi

Alloy C22

246 bar
3568 psi

218 bar
3162 psi

185 bar
2683 psi

154 bar
2234 psi

Super Duplex *
Low Pressure

842 bar
12212 psi

783 bar
11357 psi

702 bar
10182 psi

 

Super Duplex *
High Pressure

1375 bar
19943 psi

1206 bar
17492 psi

1090 bar
15809 psi

 

Tantalum

81 bar
1175 psi

63 bar
914 psi

51 bar
740 psi

 

* Minimum operating temperature for Super Duplex stainless steel is -40 °C. Maximum pressure may be lower depending on the design type.

Uncertainties and flow measurement turn-down
Uncertainties and flow measurement turn-down

The turn down capability from Qₙₒₘ of the flow sensor is driven mainly by its zero point stability. At the very low end of the measuring range the uncertainty (u) is dominated by zero point stability.

• Zero point stability of a standard sensor with general accuracy calibration: 0.0039 kg/min (0.0086 lbs/min).
• Zero point stability of a premium sensor with high accuracy or focused calibration: 0.0016 kg/min (0.0035 lbs/min).
• uncertainty = calibration uncertainty, if: flow ≥ (zero point stability / (base uncertainty/100))
• uncertainty = (zero point stability/Q)*100, if: flow < (zero point stability / (base uncertainty/100))

Uncertainties from environmental and process conditions
Uncertainties from environmental and process conditions

If sensors are not zeroed at operating conditions, minor additional uncertainties can arise from elevated temperatures and pressures:

• ±0.00053% of maximum flow per °C
• ±0.00042% of maximum flow per bar

Process temperature effect on density:
• Additional uncertainty of ±0.00013 g/cm³ per °C difference from calibration temperature with standard density calibration.
• Additional uncertainty of ±0.00011 g/cm³ per °C difference from calibration temperature with enhanced density calibration.
• This effect can be mitigated by a simple field density adjustment at operating conditions.

Process pressure effect on mass flow:
• The effect of pressure on flow measurement is 0.0001% of rate per bar. Compensation is possible by pressure sensor input (analog input or digital write) or manual value entry into the transmitter.

Process pressure effect on density:
• The effect of pressure on density measurement is 0.000075 g/cm³ per bar. Compensation is possible by pressure sensor input (analog input or digital write) or manual value entry into the transmitter.

Applying Premium calibration to higher pressure, special materials and/or non-standard temperature models may show higher zero uncertainties (up to 3 times higher than standard sensor).

Couldn’t find what you were looking for?

Contact us with your specification for assessment and availability of other sensor configurations.

Other wetted materials may be possible for chemical compatibility, lower pressure drop, abrasion allowance or other application specific requirements. Rheonik can provide nearly any material for the wetted parts.